Keysight (Agilent HP) 8593A Repair Service

8593A spectrum analyzer 9kHz-22GHz
Need your HP 8593A Spectrum Analyzer repaired or calibrated? Our veteran-owned lab handles both, with free estimates and quick turnaround.
The Agilent 8593A is an easy-to-use microwave spectrum analyzer that offers a wide range of performance, features, and optional capability to meet your measurement needs. Downloadable measurement personalities combine with optional plug-in performance to provide tailored solutions for your application.
8593a
Any failures or unforeseen issues identified during the calibration phase will be detailed in an estimate submitted for your approval prior to any further action. Repair charges are determined by the actual time and materials required. Please be aware that customers are responsible for all associated shipping costs, to and from our location. This initial estimate is provided free of charge.
8593A Specifications – click to expand

8593A Specifications

Displayed average noise level:.

SpecificationValue
Frequency range9 kHz to 22 GHz, 26.5 GHz (Opt 026)
Span range0 Hz (zero span), [10 x N] kHz to 19.25 GHz
Span accuracySpan 10 MHz: ±3%
Sweeptime RangeSpan = 0 Hz or > 10 kHz: 20 ms to 100 s
Sweeptime Accuracy20 ms to 100 s: ±3%
Amplitude range-114 to +30 dBm
8593A Options – click to expand

8593A Options

  • 004: Precision Frequency Reference
  • 010: 50 ohm Built-in Tracking Generator
  • 021: HP-IB Interface
  • 023: RS-232 Interface
  • 026: 26.5 GHz Frequency Extension
  • 101: Fast Time Domain Sweeps
  • 102: AM/FM Demodulator and TV Sync Trigger
  • 103: Quasi-Peak Detector/AM-FM Demodulator
  • 105: Time-gated Spectrum Analysis
HP 8593A Error Messages – click to expand

HP 8593A Error Messages

The HP 8593A spectrum analyzer (9 kHz to 22 GHz) shows text messages on screen when something is wrong. Each is marked (H) hardware, (U) user/setup, or (M) information.

First step: connect CAL OUT to the input and run CAL FREQ & AMPTD and CAL YTF, then press CAL STORE. A hardware (H) message that comes back after a good self-cal means the 8593A needs repair.

8593A notes

  • The 8593A has a YIG-tuned filter; run CAL YTF after CAL FREQ & AMPTD with the 100 MHz comb output cabled to the input.
  • The 8593A has counter-lock hardware for better frequency accuracy.
  • SRQ numbers on the 8593A screen are service requests: SRQ 110 means hardware broken, SRQ 140 means an illegal command.

8593A hardware messages

MessageTypeMeaning
FAIL: xxxx xxxxxxxxxxHA circuit fault was found during the power-up self-test. The 4-digit and 10-digit code identifies the failed circuit (see the service manual).
FREQ UNCALU/HYTO tuning failure. Common when default correction factors are in use; run CAL FREQ. It also flashes briefly during CAL FREQ or when turning the knob, which is normal.
CAL SIGNAL NOT FOUNDU/HThe CAL OUT signal could not be found. Check the cable from CAL OUT to the input. If it is connected, enter the −37 Hz passcode (FREQUENCY, −37, Hz) before CAL FREQ or CAL FREQ & AMPTD.
CAL: GAIN FAILHSignal level too low during CAL AMPTD.
CAL: SPAN SENS FAILHThe span sensitivity calibration failed.
CAL: FM SPAN SENS FAILHCould not set the FM coil span sensitivity.
CAL: RES BW AMPL FAILHThe relative loss of a resolution bandwidth filter is wrong.
CAL: DATA NOT STORED CAL AMP NEEDEDU/HCorrection factors are corrupt and cannot be stored. Run CAL FREQ & AMPTD, then CAL STORE.
CAL: cannot execute CALAMP enter: 0 dB PREAMP GAINU/HSet EXT PREAMP gain to 0 dB before running CAL AMPTD.
CONF TEST FAILHThe confidence test failed.
POS-PK FAILHPositive-peak detector failed.
SAMPLE FAILHSample detector failed.
VID-BW FAILHVideo bandwidth check failed.
RES-BW NOISE FAILHNoise floor too high at the bandwidth shown.
RES-BW SHAPE FAILHA resolution bandwidth 3 dB width is out of specification.
STEP GAIN ATTEN FAILHIF step gain failed.
COMB SIGNAL NOT FOUNDU/HThe 100 MHz comb signal was not found. Check that COMB OUT is cabled to the input.
CAL: LOST COMB SIGNALU/HComb signal too weak to finish CAL YTF. Use a short, low-loss SMA-to-Type-N cable from the comb output to the input.
ADC-GND FAILHThe processor board failed its analog-to-digital converter ground check.
ADC-TIME FAILHThe processor board failed its analog-to-digital converter timing check.
ADC-2V FAILHThe processor board failed its analog-to-digital converter 2 V reference check.

8593A calibration, setup and information messages

MessageTypeMeaning
CAL: USING DEFAULT DATAMCal data is corrupt, so default correction factors are in use. An interrupted self-cal can cause this.
MEAS UNCALUSweep time, span and bandwidth settings make the measurement uncalibrated.
CAL: PASSCODE NEEDEDMThe function needs the service passcode (center frequency −37 Hz).
SRQ 110MHardware broken service request. Look for the hardware message shown with it.
SRQ 140MIllegal command was received over the bus.
INVALID SYMTAB ENTRY: SYMTAB OVERFLOWUToo many user-defined items or downloadable programs in memory; delete some with DISPOSE USER MEM. It can also mean a failed processor board.

8593A remote programming and memory card messages (U)

These come from bad remote commands, memory card problems or bus setup, not hardware: COMMAND ERROR, PARAMETER ERROR, INVALID ITEM, INVALID VALUE PARAMETER, INVALID ENTER FORMAT, INVALID OUTPUT FORMAT, INVALID BLOCK FORMAT: IF STATEMENT, INVALID COMPARE OPERATOR, INVALID REPEAT MEM OVFL, INVALID REPEAT NEST LEVEL, INVALID ACTDEF, INVALID VARDEF, INVALID KEYNAME, INVALID AUNITS, INVALID DETECTOR, INVALID TRIGGER MODE, INVALID TRACE, INVALID TRACE NAME, INVALID REGISTER NUMBER, INVALID SAVEREG, INVALID STORE DEST, INVALID CHECKSUM: USTATE, INVALID AMPCOR: FREQ, INVALID WINDOW TYPE, CONFLICT TABLE OVERFLOW, CAL: NO YTO AVAILABLE, SOFTKEY OVFL, UNDEF KEY, TABLE FULL, INVALID HP-IB ADDRESS/OPERATION, INVALID HP-IB OPERATION REN TRUE, INVALID RS-232 ADDRESS/OPERATION, INVALID CARD, INVALID CARD: DIRECTORY / NO CARD / TYPE, INVALID FILE: NO ROOM, NO CARD FOUND.

Hardware message on your 8593A? Analyzer Repair Inc. repairs and calibrates the HP 8593A in-house in Cocoa, Florida, even though HP support ended long ago. Tell us the message you see; estimates are free once the unit is in our lab. Contact us for service.

Message list based on the HP 8590B/8592B Installation, Verification and Operation Manual, Appendix A.

Located in: Spectrum Analyzers